MICROMVL Image Release History |
|
|
|
|
|
|
Version |
Release Date |
|
Changes |
|
|
|
|
|
|
6.11.23 |
5/26/2021 |
|
1. Supported Different Ranges of Foxboro Sensor |
|
|
|
|
|
|
6.11.22 |
3/23/2021 |
|
1. Corrected unstable RTD Input Issue |
|
|
|
|
|
|
6.11.21 |
1/21/2021 |
|
1. Added Foxboro Transmitter
Interface |
|
|
|
|
2. Reset Turbine Offset at Startup |
|
|
|
|
3. Added ADC Counts |
|
|
|
|
|
|
6.11.20 |
5/30/2018 |
|
1. Increased Timer for A/D
Converter (Analog Inputs) |
|
|
|
|
|
|
6.11.19 |
6/6/2017 |
|
1. Reduced Time for A/D Converter |
|
|
|
|
2. Corrected Single Stream Meter#2 Flow Directional Flag |
|
|
|
|
|
|
6.11.18 |
2/27/2014 |
|
1. Added Floating Point Mode |
|
|
|
|
2. Added Modbus Shift - Floating Point |
|
|
|
|
|
|
6.11.16 |
10/21/2013 |
|
1. Added Master Port Write |
|
|
|
|
|
|
6.11.15 |
9/9/2013 |
|
1. Corrected Analog Input#3
Reading |
|
|
|
|
|
|
6.11.14 |
7/31/2013 |
|
1. Corrected Version 6.11.13 for
the 2nd 485 Port Communication |
|
|
|
|
|
|
6.11.13 |
5/13/2013 |
|
1. Set Less Sensitivity Default
Value for the Display Keys |
|
|
|
|
2. Modified Display Routine. |
|
|
|
|
|
|
6.11.12 |
3/15/2013 |
|
1. Modified Communication Routine |
|
|
|
|
2. Reduced Sdevice Timer |
|
|
|
|
3. Set Display Contrast Initialization to 35 |
|
|
|
|
4. Set Double Write to Download Mode via RS232 Port |
|
|
|
|
|
|
6.11.11 |
2/24/2012 |
|
1. Correction - Master-Slave
Activation for Micro2009 and Later Model |
|
|
|
|
|
|
6.11.10 |
10/14/2011 |
|
1. Correction - New23/24 with
Live Densitometer Temperature and Densitometer Pressure |
|
|
|
|
|
|
6.11.09 |
11/4/2010 |
|
1. Correction - Calibration Mode
- AGA7 |
|
|
|
|
|
|
6.11.08 |
7/29/2010 |
|
1. Added Forward/Reverse Flow
Rate Modbus Registers |
|
|
|
|
|
|
6.11.07 |
3/31/2010 |
|
1. Modified Reset Routine |
|
|
|
|
2. Correction - Densitometer Pressure mA Reading |
|
|
|
|
|
|
6.11.06 |
1/28/2010 |
|
1. Added Rosemount Meter Type |
|
|
|
|
|
|
6.11.05 |
10/16/2009 |
|
1. Correction - VenCone Equation |
|
|
|
|
|
|
6.11.04 |
9/16/2009 |
|
1. Modified Display
Initialization |
|
|
|
|
|
|
6.11.03 |
8/28/2009 |
|
1. Correction - Table 12,13,15,17
Density@60 |
|
|
|
|
|
|
6.11.02 |
8/1/2009 |
|
1. Set RAM/Flash to One Wait
State |
|
|
|
|
|
|
6.11.01 |
3/27/2009 |
|
1. Corrected 6.11.0 for
Supporting Old Hardware Units |
|
|
|
|
|
|
6.11.0 |
3/20/2009 |
|
1. Added Support for Old and New
Hardware Units |
|
|
|
|
|
|
6.10.9 |
12/12/2008 |
|
1. Added Vencone Flow Equation |
|
|
|
|
|
|
6.10.08 |
11/12/2008 |
|
1. Added Flow Direction Override thru. Modbus Option |
|
|
|
|
2. Added Batch End Modbus Registsers |
|
|
|
|
3. Correction - Slave Density Reading |
|
|
|
|
|
|
6.10.07 |
5/20/2008 |
|
1. Add Multivar DP Flow Direction
Indication Delay |
|
|
|
|
|
|
6.10.06 |
5/2/2008 |
|
1. Correction - NIST14 Fluid
Mixtures |
|
|
|
|
2. Expand Range of NIST14 Calculation |
|
|
|
|
|
|
6.10.05 |
3/11/2008 |
|
1. New Rosemount Average Routine |
|
|
|
|
2. Modify Audit Trail Report Routine |
|
|
|
|
3. Add Reverse Hour/Daily Totalizer |
|
|
|
|
|
|
6.10.04 |
1/15/2008 |
|
1. Correction -
API2565-Proplene/Ethylene for Version 6.10-6.10.3 |
|
|
|
|
|
|
6.10.03 |
1/9/2008 |
|
1. Modify GPA-TP15(2007) |
|
|
|
|
|
|
6.10.02 |
10/4/2007 |
|
1. Add Yesterday Day and Last Hour Modbus Registers |
|
|
|
|
|
|
6.10.01 |
5/25/2007 |
|
1. Diagnostic Screen - Display
Density in GM/CC |
|
|
|
|
|
|
6.10.00 |
3/20/2007 |
|
1. Add - API 2004
Crude/Refined/Special Product Tables(Replacements for 5/6,23/24). |
|
|
|
|
2. Correction - New Table 23/24 (Replacements for OLD
Table 23/24) |
|
|
|
|
|
|
6.09.14 |
1/8/2007 |
|
1. Modify Display Sensitivity Key
Routine |
|
|
|
|
|
|
6.09.12 |
6/30/2006 |
|
1. Add - Apply Density Correction
Factor to mA Density |
|
|
|
|
2. Correction - Gravity Out of Range Volume
Re-Calculation for a New Product |
|
|
|
|
|
|
6.09.11 |
2/3/2006 |
|
1. The Size of Historical Daily Data is Increased to 35 |
|
|
|
|
|
|
6.09.10 |
1/10/2006 |
|
1. Add Digital Input/Output #4
Option |
|
|
|
|
|
|
6.09.09 |
10/4/2005 |
|
1. Correction - Analog Input 1-4
Reading (Version 6.09.08) |
|
|
|
|
|
|
6.09.08 |
8/23/2005 |
|
1. Add Analog Input Expansion Option |
|
|
|
|
|
|
6.09.07 |
7/19/2005 |
|
1. Correction - Station Display
Unit Text |
|
|
|
|
|
|
6.09.6 |
3/31/2005 |
|
1. Correction - Flow Rate Scale
Data |
|
|
|
|
2. Increase Range of NIST14 Density Calculation |
|
|
|
|
|
|
6.09.5 |
2/22/2005 |
|
1. Add - Daylight Saving Time
Option |
|
|
|
|
2. Correction -Save Daily Total on Daily Batch Report |
|
|
|
|
|
|
6.09.4 |
1/10/2005 |
|
1. Add - Batch Report and Daily
Report Show Cumulative Total |
|
|
|
|
2. Add - Disable Interval Report If Threre Is No Flow
During The Batch. |
|
|
|
|
|
|
6.09.3 |
12/22/2004 |
|
1.Correction - PPMIX Using Live
Densitometer |
|
|
|
|
|
|
6.09.2 |
11/24/2004 |
|
1. Modify Modbus Shift Routine |
|
|
|
|
2. Modify Boolean Statement/Program Variable Statement
Routine |
|
|
|
|
3. Add Battery Voltage Reading |
|
|
|
|
4. Enable/Disable Front Panel Scan Key |
|
|
|
|
5. Enable/Disable Display Backlight Function |
|
|
|
|
6. Correction - Venturi Y Factor |
|
|
|
|
7. Add Calibration Parameters to be Part of the
Configuration Software |
|
|
|
|
|
|
6.09.1 |
8/26/2004 |
|
1. Add - V-Cone Flow Equation |
|
|
|
|
|
|
6.09.0 |
8/4/2004 |
|
1. Add - TPU Frequency Counters |
|
|
|
|
2. Correction - Modbus ASCII Mode |
|
|
|
|
|
|
6.08.9 |
5/20/2004 |
|
1. Add - Net and Mass Cumulative
Modbus Registers |
|
|
|
|
2. Correction - Broadcast Command |
|
|
|
|
|
|
6.08.8 |
10/27/2003 |
|
1. Set RAM/ROM Wait State to 1 |
|
|
|
|
|
|
6.08.7 |
9/26/2003 |
|
1. Modify PPMIX Range |
|
|
|
|
2. Add Flow Direction Override Modus Register |
|
|
|
|
3. Increase Stack |
|
|
|
|
4. Correction - 16
Bits Modbus Shfit Offset |
|
|
|
|
|
|
6.08.6 |
8/26/2003 |
|
1. Add - Change Flow Direction
Through Modbus |
|
|
|
|
|
|
6.08.5 |
8/7/2003 |
|
1. Improve Multi.Variable
Interface Routine |
|
|
|
|
|
|
6.08.4 |
7/21/2003 |
|
1. Add - WatchDog Timer
(3022) |
|
|
|
|
|
|
6.08.3 |
7/9/2003 |
|
1. DP direction indication is
based on 5 seconds averaged dp |
|
|
|
|
|
|
6.08.2 |
6/10/2003 |
|
1. Modify Audit Trail Report -
Add Net and Mass Cumulative Total |
|
|
|
|
|
|
6.08.1 |
1/15/2003 |
|
1. Increase Modbus Shift Range
from 5001 to 5119 |
|
|
|
|
|
|
6.08.0 |
11/8/2002 |
|
1. Add Status Input Assignments -
Display Toggle/Display Freeze |
|
|
|
|
2. Add 5071-5079 Program Variables (1 Decimal Inferred) |
|
|
|
|
3. Add New Modbus Registers - Gross, Net, Mass Volume
(One Second) |
|
|
|
|
|
|
6.07.9 |
10/31/2002 |
|
1. Correction - New 23/24 Table Iteration Routine (Using
Live Densitometer) |
|
|
|
|
|
|
6.07.8 |
10/30/2002 |
|
1. Correction - New 23/24 Table Iteration Routine (Using
Live Densitometer) |
|
|
|
|
|
|
6.07.7 |
10/11/2002 |
|
1. Correction - Analog Input Routine |
|
|
|
|
|
|
6.07.6 |
9/6/2002 |
|
1. Increase DP Range of Multi-Variable to 1075 |
|
|
|
|
|
|
6.07.5 |
8/3/2002 |
|
1. Correction - Use Densitometer Pressure, not Meter
Pressure |
|
|
|
|
|
|
6.07.4 |
6/3/2002 |
|
1. Correction - Multi-Variable DP Calibration |
|
|
|
|
2. Correction - Analog Input Calibration |
|
|
|
|
|
|
6.07.3 |
3/18/2002 |
|
1. Add New Last Month Report |
|
|
|
|
2. Delete Batch Schedule Routine |
|
|
|
|
|
|
6.07.2 |
2/18/2002 |
|
1. Add New Modbus Registers - Digital In/Output Bits |
|
|
|
|
|
|
6.07.1 |
1/3/2002 |
|
1. Set Wait State to 2 |
|
|
|
|
|
|
6.07.0 |
12/20/2001 |
|
1. Add Linear Meter Factor Routine (Reverse Direction) |
|
|
|
|
|
|
6.06.9 |
12/11/2001 |
|
1. Add New Modbus Registers - Forward/Reverse Lineaer
Meter Factor |
|
|
|
|
|
|
6.06.8 |
11/28/2001 |
|
1. Correction - Table 24C - Alpha T Calculation |
|
|
|
|
|
|
6.06.7 |
11/10/2001 |
|
1. Add New Modbus Registers - Last Batch Data |
|
|
|
|
|
|
6.06.6 |
10/23/2001 |
|
1. Correction - RTD Calibration |
|
|
|
|
|
|
6.06.5 |
10/2/2001 |
|
1. Increase Baud Rate to 19200 |
|
|
|
|
|
|
6.06.4 |
9/15/2001 |
|
1. Add Mass Pulse Selection |
|
|
|
|
|
|
6.06.3 |
9/10/2001 |
|
1. Add Boolean Logic To Do Remote Control |
|
|
|
|
|
|
6.06.2 |
8/17/2001 |
|
1. Correction - Analog Output Maximum |
|
|
|
|
|
|
6.06.1 |
8/8/2001 |
|
1. Revise Communication Routine |
|
|
|
|
|
|
6.06.0 |
7/16/2001 |
|
1. Add New Modbus Registers - Yesterday and Last Batch
Data |
|
|
|
|
|
|
6.05.5 |
7/9/2001 |
|
1. Correction - Analog Output Drop |
|
|
|
|
2. Correction - Slave Alarm Message |
|
|
|
|
3. Add New Modbus Registers for Previous Batch Data |
|
|
|
|
|
|
6.05.4 |
7/1/2001 |
|
1. Correction - Pulse Output |
|
|
|
|
2. Correction - Increase RTXC Stack |
|
|
|
|
|
|
6.05.3 |
6/10/2001 |
|
1. Correction - RTD Reading (Increase RTD Gain) |
|
|
|
|
|
|
6.05.2 |
4/20/2001 |
|
1. Correction - Densitometer Alarms |
|
|
|
|
|
|
6.05.1 |
4/15/2001 |
|
1. Correction - Analog Output, PID Alarms |
|
|
|
|
|
|
6.05.0 |
4/12/2001 |
|
1. Correction - UGC2 Calculation |
|
|
|
|
2. Add Analog Output #1-#4 |
|
|
|
|
|
|
6.04.0 |
2/11/2001 |
|
1. Correction - Status Input/Switch Output #1/#4 Setting |
|
|
|
|
|
|
6.03.1 |
12/22/2000 |
|
1. Add RTD Input |
|
|
|
|
2. Add PPMIX |
|
|
|
|
|
|
6.03.0 |
11/4/2000 |
|
1. Revise RS-485 Routine |
|
|
|
|
|
|
6.02.0 |
10/10/2000 |
|
1. Correction - Switch Output, Spare Data Alarms |
|
|
|
|
|
|
6.01.0 |
10/5/2000 |
|
1. Correction - Density
Calculation |
|
|
|
|
2. Set Data Ready Flag After Retrieving Historical Data |
|
|
|
|
|
|
|
|
|
|